Introduction
The precision of
laboratory results is important to help clinicians to diagnose and to monitor
the patient treatment. The automated analyzer is used for analysis of the
patient samples for routine chemistry and urine tests, immunology and
immunoassay. The instrument must be precise to deliver the patient results.
This is very important in patient management.
METHODOLOGY
PRELIMANARI PRECISION
·
Level 1 and level 2 quality control recommended by
manufacturer were run triplicate per day for five day.
·
The mean, variance, standard deviation and coefficient
variation experiment were calculated for repeatable and intermediate.
·
The repeatable and intermediate coefficient variation were
compared with the repeatable and intermediate coefficient variation
manufacturer followed by The Clinical Laboratory Standard Institute (CLSI EP
10).
If unacceptable for both or
other one perform ,verification precision
VERIFICATION
PRECISION
intermediate manufacturer(TSDmnf) were
calculated based on mean experiment and
coefficient variation manufacturer.
·
The standard deviation for repeatable and/or
intermediated manufacturer were compared with the standard deviation experiment
(repeatable and/or intermediate where applicable) followed by The Clinical
Laboratory Standard Institute (CLSI EP 15)
·
If unacceptable, the troubleshooting were perform and
repeat the preliminary precision
Results
The achievement of the analyzer for routine chemistry
and urine tests, immunology and immunoassay for the level 1 QC based on the coefficient
variation manufacturer claim were 54% for repeatable and 78% for intermediate
at the preliminary precision. 46% of the level 1 QC for repeatable and 22
% of the intermediate were acceptable at
the verification precision.
For the level 2 QC results, the performance based on
the coefficient variation of manufacturer claim is 56% and 77% for
intermediate. Therefore, 44% of
the;level; 2 QC for repeatable and 23% of the intermediate were acceptable
based on the verification precision.
Conclusion
Based on the studies, all the results of the routine
chemistry and urine tests, immunology
and immunoassay are precised for patient management as long as maintenance
schedule, the procedure of the pre-analytical as sample collection,
transportation, time receiving and processing of samples are followed.
No comments:
Post a Comment