Monday, July 29, 2019
Tuesday, February 13, 2018
Tuesday, April 18, 2017
Monday, April 10, 2017
Friday, April 7, 2017
Thursday, April 6, 2017
Wednesday, April 5, 2017
Monday, April 3, 2017
METHOD VERIFICATION
PRELIMINARY STUDY: WITHIN RUN PRECISION
Method verification is a one- time process to determine performance characteristics before a
test system is utilized for patient testing. The experimental confirmation of performance
usually include preliminary study (precision check), linearity, comparison study and
reference range.
Preliminary Study
The preliminary study is applied only to newly installed instruments or new measurement
procedure in the laboratory. This is very important to verify the precision performance of
method in the insert kit is fulfilled before the new method is utilize for patient sample testing.
The basic statistics calculation as below will used to verified the precision of the new
method:
· Mean = ( Σ x)/ n , n= number of observation
· Variance, σ2 = Σ (Xi – Mean )2 / (n-1)
· Standard Deviation, SD= √ σ2
· Coefficient Variation ( CV ) = (SD/mean) x 100.
The main objective of preliminary study is to verify the precision performances in the insert
kit are fulfilled before we use the new method in our laboratory. These include CV within
run precision and CV intermediate or total precision.
We will use control material at 2 or 3 levels as recommended by the manufacturer to asses precision within run and total precision for verification of manufacturer’s claim. We will
accept the manufacturer precision within run and total precision if precision experiment
within run and total precision is less than manufacturer claim.
The experiment procedures for the preliminary study is to run triplicate per day within two
hours for five days using control material as recommended by the manufacturer. We
calculate the daily mean and variance (σ2) analytical. From daily mean and variance
analytical, we calculate overall mean, standard deviation (SD) and coefficient variation (CV) analytical for within runs, between days and total CV. Next, we compare CV analytical withinrun and total CV with manufacturer claims. If CV analytical within run and total CV analytical is less than CV within run and total precision manufacturer claim, we will accept the
precision manufacturer claims. If more, we will continue to calculate SD verification (SDV)
from CV manufacturer, chai- distribution and degree of freedom. We only accept if SD
analytical less than SDV.
Preliminary Study: Within Run
The procedure of the preliminary study is run all the quality control (QC ) material as
recommended by manufacturer triplicate within two hour for five day and record all the data.
Daily mean and variance (σ2) were calculate.
Example : Albumin, the total number of level QC are two.
The calculation for this preliminary study, only for level 2 QC. The mean level 2 is 38 g/l, CV within run is 0.5% and Total CV is 1.1%.In real situation, we must perform precision study
for level 1 and level 2. The data triplicate for five days with mean and variance as below:
Day
|
Run 1
|
Run 2
|
Run 3
|
Mean
|
Variance
|
1
|
37.7
|
37.5
|
37.8
|
37.73
|
0.0633
|
2
|
37.3
|
37.5
|
37.7
|
37.50
|
0.0400
|
3
|
38.4
|
37.6
|
38.1
|
38.03
|
0.1633
|
4
|
38
|
37.3
|
38.2
|
37.83
|
0.2233
|
5
|
38.1
|
37.8
|
38
|
37.97
|
0.0233
|
Note: formula for mean and variance
Mean = ( Σ x)/ n , n= number of observation or number of run
Variance, σ2 = Σ (Xi – Mean )2 / (n-1)
Overall mean and variance
Overall mean = (37.73 + 37.50+38.03+37.83+37.97)/5
= 37.81 g/l
Overall variance = (0.0633+0.0400+0.1633+0.2233+0.0233)/5
Standard deviation (SD) and coefficient variation CV)
SD = √variance
= √0.1027
CV =(SD/overall mean) x 100
=( 0.3204/37.81) x 100
= 0.8474% CV analytical
CV manufacturer is 0.5% and CV analytical is 0.8474%, That means analytical CV is more
than manufacturer CV . Next step is we calculate the SD verification (SDV)
Standard deviation verification (SDV)
SDV = {(CV manufacturer/100 x overall mean) x √C} / √V
C= Percentage point ( Ӽ2 (0.05/L, V), L= number of QC level (L=2)
V= Degree of freedom ( L*n), L= number of QC level, n = number of day (n=5)
SDV = {(1.1/100 x 37.8) x √ 20.48} /√ 10
= 0.12706
SD analytical is 0.3204, therefore the SD analytical > SDV means the precision for within
run manufacturer is not acceptable. The manufacturer must perform troubleshooting and
preliminary study for within run and re-evaluate.
Note: The value C is obtain from chi-square table or from window excel
( Ӽ2 (0.025,10).
Friday, March 31, 2017
APPLICATION OF MEASUREMENT UNCERTAINTY IN CLINICAL LABORATORY
Sairi Satari
Chemical Pathology Unit
Penang Hospital
The uncertainty of measurement is a fundamental concept indicating the range within the “true value of a measurement should lie”. Although not commonly reported with the result, the calculation of measurement uncertainty (MU) has become common in the clinical laboratory routine. MU estimates are usually used in analytical troubleshooting and interpretation of pathology results. In analytical troubleshooting, the expanded uncertainty will be obtained by multiplying MU with 1.96, In the interpretation of numerical pathology results, the patients’ results are compared with the reference interval or medical decision limit and the expanded uncertainty will be obtained by multiplying MU with 1.65. If the interpretation is done by comparing with previous result from the same patient, the expanded uncertainty will be obtained by multiplying MU with 2.77. In patients’ management, MU is significant during validation of the patients’ results for accuracy and reliability.
Sunday, September 28, 2014
INTERNAL QUALITY CONTROL (IQC) DAN
PENYELESAIAN MASALAH
SAIRI SATARI
Pegawai
Sains Kimiahayat
BAHAGIAN 2
TARGET MEAN, SD DAN CV
Target mean, SD dan CV yang diabadikan
dalam graf Levey Jennings akan digunapakai selagi lot QC yang sama digunakan
sebagai material untuk memantau kualiti ujian makmal. Oleh itu, pengiraan target mean, SD dan CV perlu memenuhi kriteria
berikut;
·
Minima
data 20 dari lot QC yang hendak digunapakai dan data membentuk taburan normal,
·
Analisa
sampel QC perlu melibatkan beberapa lot kalibrator dan pertukaran lot reagent,
·
Penglibatan
beberapa operator dalam penyediaan larutan kalibrator, QC dan analisisa,
·
Penggunaan
glass pipette
yang sesuai semasa
menyediakan larutan kalibrator dan QC.
Kriteria di atas perlu dipenuhi kerana
target mean, SD dan CV membawa maksud tersurat iaitu ralat sistematik dan ralat
random. Ralat-ralat ini dijelmakan sebagai target mean dan target SD dan
diabadikan dalam graf Levey Jennings sebagai mean ±2SD. 2SD. Ia adalah nilai
95% confident limit. Ralat yang wujud
ini perlu konsisten selagi kita menggunakan lot QC yang sama. Pertambahan ralat
atau kewujudan ralat baru menyebabkan perubahan target mean dan target SD. Oleh
itu kita perlu mengira mean, SD dan CV setiap bulan. Nilai mean setiap bulan
perlu sama dengan target mean secara statistik dan CV setiap bulan perlu kurang
dari target CV. Perubahan lot QC memerlukan kita mengira semula target mean dan
SD.
Taburan normal dapat dikenalpasti
apabila set data memberi nilai mean, median dan mode yang sama (mean=median
=mode)
![]() |
| Taburan normal |
![]() |
| Mean +/- 2 SD |
Setiap kali kita menganalisa sampel
QC, keputusan QC perlu berada dalam julat mean + 2SD atau mean -2SD dalam
keadaan seimbang. Oleh kerana kita sukar untuk menentukan keputusan QC berada
dalam julat mean + 2SD atau mean - 2SD yang seimbang, maka graf Levey Jennings
digunapakai dalam memantau QC harian. Westgard
rule pula digunakan untuk memantauan ralat sistematik dan ralat random semasa
proses analitikal.
![]() |
| Graf Levey Jenning |
![]() |
| Multi Westgard rule |
Penyediaan larutan QC dan kalibrator
perlu menggunakan glass pipette yang
sesuai. Jika kita ingin menyediakan larutan 5 ml, maka kita perlu menggunakan pipette 5ml dan bukannya pipette 10 ml. begitu juga untuk
menyediakan larutan 1ml, kita perlu menggunakan glass pipette 1 ml. Setiap pipette mempunyai uncertainty. Pipette yang mempunyai isipadu yang besar mempunyai uncertainty yang besar, begitu juga sebaliknya.
BERSAMBUNG
Monday, September 22, 2014
INTERNAL QUALITY CONTROL
(IQC) AND PROBLEM SOLVING
SAIRI
SATARI
Biochemist
Part 1
Part 1
RECTIFICATION IS
ESSENTIAL, ALBEIT ROUTINE PROCEDURES
Internal
Quality Control (IQC) is a process or system to monitor the quality of
laboratory testing in terms of accuracy and precision. It includes pre-analytical
and analytical phase. IQC should be able to detect the changing of systematic error
and random error during the analytical process. Therefore, when we determine
the target mean and SD or CV, the systematic error and random error should be
taken into account. Drastic changes of the systematic error and random error
will be affect the mean and SD or CV.
![]() |
Target
Mean, SD and CV will be change
|
Levey
Jenning’s chart and Westgard rule are the common procedures used in the
laboratories to determine the systematic error and random error. The IQC must be compliance with the Wesgard
rule before the patient results are released.
Usually,
when we determine the target mean, SD or CV, we will collect 20 data point from
selected level of quality control (QC) within 20 working days. We calculate the mean, SD, and CV from the
data. If the calculated mean is within the manufacturer’s mean range, then the
mean and SD will be used as a target mean, and target SD. The target mean and
SD will be stated in the Levey Jennings chart as mean ±2 SD (2 SD is 95%
confident limit based on the normal distribution). The acceptability of our
daily QC will be determine by Westgard rule.
The Levey Jenning’s chart and the Wesgard rule
are based on the normal distribution. Thus, the setting of the target mean and
SD or CV based on 20 data points only are not sufficient. The setting of target
mean and SD or CV should take into account that the data collected should be in
a normal distribution. Therefore, we must determine whether
20 data points will form a normal distribution or not. If not, we need to add
more data until it forms a normal distribution.
The
target mean and SD form manufacturer recommendation are wrong in term of
setting procedure of the target mean and SD in the laboratory. We must
calculate our target mean and SD based on our condition in the laboratory.
![]() |
| Levey Jenning Chart |
When
we perform repeatable analysis to the same QC lots, the data collected will
generally be in a normal distribution but, we still have to prove that the data
is in the normal distribution by ensuring that the mean, median and mode generate
the same value. The target mean will reflect the systematic error and SD or CV
will reflects the random error. Both of these errors will have to be consistent
as long as we are using the same lot of QC in our analysis. Thus, the setting
of the target mean and SD QC need to be based on the normal distribution at 95%
confident limit or Mean ± 2SD. If the mean setting and SD is not based on the
normal distribution, then we will have difficulty in solving QC non-compliance during
analysis.
![]() |
The Levey Jenning’s Chart and the Westgard
Rule are based on the Normal Distribution and it can be help us to solve the
problem for non-conformant daily QC.
|
TO BE CONTINUED
INTERNAL QUALITY CONTROL (IQC) DAN
PENYELESAIAN MASALAH
SAIRI SATARI
BAHAGIAN 1
“YANG
BIASA TIDAK SEMESTINYA BETUL. PERBETULKAN YANG BIASA”
Internal Quality Control (IQC) adalah suatu
proses atau sistem untuk memantau kualiti ujian makmal dari aspek ketepatan
(accuracy) dan kejituan (precision). Ia merangkumi fasa pre-analitikal dan
analitikal. !QC sewajarnya dapat mengesan perubahan ralat sistematik dan ralat
random semasa proses analitikal. Sehubungan itu, apabila kita menentukan target
mean dan SD atau CV, penyumbang ralat sistematik dan random perlu diambil kira.
Perubahan yang drastik terhadap ralat sistematik dan ralat random akan mengubah
mean dan SD atau CV.
Graf Levey Jennings dan Westgard rule adalah prosedur biasa yang digunapakai di makmal dalam menentukan kepatuhan IQC sebelun keputusan makmal dikeluarkan. Adalah menjadi kebiasaan, apabila kita menentukan target mean, SD atau CV, kita hanya mengumpulkan 20 data dalam jangka masa 20 hari bekerja dari lot IQC yang hendak digunakan. Dari 20 data ini kita kira mean, SD dan CV. Apabila mean yang dikira terkandung dalam julat manufacturer, maka mean dan SD tersebut digunapakai sebagai target mean dan target SD dan diabadikan dalam graf Levey Jennings. Wesgard rule digunakan bagi menentukan IQC harian kita sama ada dapat diterima atau sebaliknya.
![]() |
| Jitu dan tepat |
![]() |
| Jitu tetapi tidak tepat |
Graf Levey Jennings dan Westgard rule adalah prosedur biasa yang digunapakai di makmal dalam menentukan kepatuhan IQC sebelun keputusan makmal dikeluarkan. Adalah menjadi kebiasaan, apabila kita menentukan target mean, SD atau CV, kita hanya mengumpulkan 20 data dalam jangka masa 20 hari bekerja dari lot IQC yang hendak digunakan. Dari 20 data ini kita kira mean, SD dan CV. Apabila mean yang dikira terkandung dalam julat manufacturer, maka mean dan SD tersebut digunapakai sebagai target mean dan target SD dan diabadikan dalam graf Levey Jennings. Wesgard rule digunakan bagi menentukan IQC harian kita sama ada dapat diterima atau sebaliknya.
Graf
Levey Jennings dan Wesgard rule
adalah berasaskan taburan normal. Oleh itu, Penetapan target mean dan SD atau
CV hanya berdasarkan 20 data semata-mata adalah tidak mencukupi.Penetapan
target mean dan SD atau CV perlu mengambil kira bahawa, data yang dikumpulkan
perlu berada dalam taburan normal. Oleh itu, kita perlu menentukan samada 20
data yang dikumpulkan itu membentuk taburan normal atau pun tidak. Jika tidak,
kita perlu menambah data sehingga membentuk taburan normal.
Amalan
menggunakan target mean dan SD dari manufacturer
adalah salah dari segi prosedur penetapan target mean dan SD IQC di makmal.m
Justeru, kita perlu mengira target mean dan SD makmal kita.
![]() |
| Graf Levey Jenning |
Apabila kita melakukan analisa
berulang kali kepada IQC yang sama lot, data yang dikumpulkan secara umumnya akan
membentuk taburan normal. Walaupun begitu, kita perlu membuktikan bahawa data
tersebut membentuk taburan normal dengan memastikan bahawa mean ,median dan mod
memberi nilai yang sama. Target mean adalah mengambarkan ralat sistematik dan
SD atau CV mengambarkan ralat random. Kedua-dua ralat ini perlu konsisten sepanjang
masa selagi kita menggunakan lot yang sama dalam analisis kita. Justeru,
penetapan standard mean dan SD IQC perlu berdasarkan taburan normal data pada
95% confident limit atau Mean ± 2SD.
Jika penetapan mean dan SD tidak berasaskan taburan normal, maka kita akan
mengalami kesukaran dalam menyelesaikan masalan ketidakpatuhan IQC semasa
menjalankan analisa.
![]() |
| 95% confident limit (tburan normal)
Berdasarkan taburan normal ini maka Graf
Levey Jennings dan Westgard rule dapat
membantu dalam menyelesaikan masalah ketidakpatuhan IQC harian.
|
BERSAMBUNG…
Sunday, August 3, 2014
LONG TERM IMPRECISION: HOW TO CALCULATE
Introduction
The
monitoring of long term laboratory imprecision is very important for patient
management, especially in the follow up for the response to treatment based on
the laboratory data. Currently, the long term imprecision is based on the
combination of the intermediate CV as CV= √ (CV12+CV22+
CV32 +….). This formula provides bigger CV compared to the generated CV by the analyzer
for the same duration.
The
aim of this study is to determine alternative method in calculation for the
long term imprecision compared to the generated CV in the same duration.
Method
The Albumin quality control data,
Level 1 and Level 2 on the Architect c8000 analyzer for six months ( July,
August, September, October, November and December 2012) were used in this study. The analyzer
generated the mean, SD and CV for the last six months duration and on monthly
basis (July to December respectively). The monthly mean, SD and CV were entered
in the excel work sheet for calculation of the long term imprecision in two
methods:
i.
Based
on the combination of monthly CV and
ii.
Based
on the combination of the monthly variance.
The value of the total CV obtained
in the above method was compared to the generated CV for the last six month
(July to December).
Results
The results from the experiment are
as the following :
1.
Automated generated Total CV by the analyzer for the duration of
July to December
|
LEVEL 1
|
LEVEL 2
|
|||||||
|
N
|
MEAN (g/l)
|
SD
|
% CV
|
N
|
MEAN ( g/l)
|
SD
|
% CV
|
|
|
July to Dec
|
231
|
40.54
|
0.827
|
2.04
|
234
|
24.7
|
0.61
|
2.46
|
|
MONTH
|
N
|
LEVEL 1
|
LEVEL 2
|
||||
|
MEAN (g/l)
|
SD
|
CV
|
MEAN
|
SD
|
CV
|
||
|
JULY
|
2
|
41
|
0
|
0.00%
|
24.5
|
0.707
|
2.89%
|
|
AUGUST
|
57
|
40.54
|
0.569
|
1.40%
|
24.745
|
0.440
|
1.78%
|
|
SEPT
|
57
|
39.72
|
0.526
|
1.32%
|
24.18
|
0.563
|
2.33%
|
|
OCT
|
54
|
40.333
|
0.476
|
1.18%
|
24.91
|
0.446
|
1.79%
|
|
NOV
|
52
|
41.481
|
0.504
|
1.22%
|
25.245
|
0.434
|
1.72%
|
|
DEC
|
9
|
41.556
|
0.527
|
1.27%
|
25.111
|
0.333
|
1.33%
|
|
Imprecision
calculation (six month)
|
Level 1 QC
|
Level 2
QC
|
|
Based on the monthly CV
|
2.86%
|
4.99%
|
|
Based on the monthly Variance
|
2.04%
|
2.46%
|
|
Automated generated by analyzer (six month)
|
2.04%
|
2.46%
|
The results showed that CV
generated by automated analyser based on the QC data July to December were
2.04% for level 1 and 2.46% for level 2. The CV calculated based on the
combination of monthly CV from July to December on the same QC data were 2.86%
for level 1 and 4.99% for level 2. On the other hand, when the CV is calculated
based on the combination of monthly variance on the same QC data, it was found
that CV for level 1 was 2.04% and level 2 was 2.46%.
Conclusion
Based
on this study, it was found that the long term CV calculated based on the
combination monthly variance gave
comparable result to that of automatically generated CV by the analyser where as combination of the
monthly CV gave much bigger CV. Therefore, it can be concluded that the CV
calculated by monthly variance is more reliable compared to the combination of
the monthly CV.
Subscribe to:
Posts (Atom)









